back Nikon Optiphot 2 instructionsCompared to first gen, Optiphot 2 differs by style, CFN optics and 100W illuminator.Acquired for (infinity) Optiphot 150 vertical illuminator and CF Plan objectives, then Optiphot 100 DIC was obtained.. Nikon Optiphot-2 MicroscopeeBay item number: 136159163096Condition: For parts or not working DescriptionPower Tested for basic functionality, does not come with additional accessories- wear and tear is visible, binocular rubber needs to be replaced - Please see photos, message with questions!! R2V3 Rating: F3 - Key Functions Working C4 - Used Good
delivered 22 Jul 2025
Condenser focus pinion gear was split. musical condenser focus gearsWhile Labophot and Optiphot condenser focus block design are similar,they are not interchangeable. However, focus pinion gear assemblies are. Replacement assemblies with metal pinion gears are available: ,
... but plenty of cannnibalized frames have intact condenser focus gear assemblies. The fine focus bushing is retained by two 1.5mm hex socket set screws. infinity head dovetailsAO/Reichert infinity microscope dovetail fits Nikon infinity: condenser centeringApprox 0.5mm-thick cardboard shims between condenser bracket and dovetailcured inability to center illumination field aperture by condenser thumbscrews;
| ||||||||||||||||||||||||
|
maintained by blekenbleu |