Optiphot 66 diascopic conversionsThe DIA-EPI base arrived 1 June 2022 with a broken condenser carrier centering screw:Seller has been contacted.. AO condenser dovetail is too large for Optiphot fork. Optiphot-66 diascopic base is essentially that of a Labophot, lacking Köhler lamp filament alignment provisions, with illumination diffuser is buried in the base, and bulb location in the lamp house not adjustable. AC power is/was switched by potentiometer on the front. Diascopic wiring in the base was hacked, originally providing 6V out 2-prong side receptacle based on adjacent switch. Someone converted that side switch to instead turn AC on/off, disconnected the side receptacle, ran AC cord from another 6V transformer thru a hole drilled in the other side rear of the base, hacked into the neutral AC line for the base transformer and added a 4 pole double throw center off switch to select either the base or external transformer, running green wire ground and unfused AC hot from the right side rear to that left side added 4PDT switch and tranformer power cord. Fine focus shaft was broken off at one end. Loosening coarse focus flanges from their threaded tubing is problematic; thread locker is wanted to set slip clutch resistance. Removing a condenser focus rack from its substage bracket seemingly requires a special wrench to loosen a pinion bearing retainer. 6V wiring to 20W base lamp was also hacked... Cutout for 6V receptacle is approx 25x11.6mm; perhaps for a small panel meter... Restored base wiringAn Optiphot-66 originally had wiring from its DIA EPI switchto adjacent receptacle for powering a (6V, 20W) EPI illuminator, which is useless without compatible cable connectors and illuminator. Wiring to that switch presumably was yellow and orange from lamp connections and red from the bridge rectifier.
This 6V system seeming could be relatively easily converted to LED; a common-emitter transistor can be nearly constant current than constant voltage... base mechanicalAbout 3mm of broken centering screw remained in the condenser carrier.After careful setup, that remaining segment was drilled out, leaving threads intact. A centering screw and field lens were still needed. Optiphot 1What was listed as a Labophot is actually an Optiphot 1 with Köhler illumination,50W lamp house and a stage that rotates (about 270 degrees) and translates. Both its condenser centering screws were unbroken. Swapping vertical illuminator's 12V bulb to 6V (LED) would enable powering by -66 6V supply... Optiphot 1, after fine focus repair, should be an overall more capable frame, except for stage size. 6 July: discopic transplantThe 1 June diascopic base has been mostly disassembled for several weeks;the seller reneged on a replacement condenser carrier and offered no partial refund. Removing substage from its focus rack requires a long (8-inch) 3mm Allen bit. Recentering an arm on a base, for illumination alignment, seems more difficult than realigning a substage on its focus rack, since condenser carriers have centering screws. Condenser rack has an adjustable spring-loaded upper stop and a removable screw for its lower stop. Sadly, diascopic lamp centering flange differ: 3- vs 4-hole mount; the 4-hole diascopic base would need centering on the epi arm with good focus. Condenser dovetail was slack, tightened by loosening right dovetail clamp screws and nudging tighter by twisting a flat blade screwdriver. Substage assembly with condenser carrier was mechanically centered in the good fine focus neck, and that neck was mechanically centered on the diascopic base. One of 4 screws securing the Optiphot-66 stage to its substage was definitely non-standard, but was reused rather than further mess with corresponding threads in the stage. 14 Sept 2022: arm transplantGiven that the 1 June DIA-EPI 66 purchase was basically a bust,and the main things wrong with 20 July DIC Optiphot 66 were: ... then swapping good arm from the original (March 2021) Optiphot 66 onto unmolested DIC Optiphot 66 base consolidates best parts. Swapping the EPI's unbroken focus into the diaphot's arm was easier than moving the condenser rack and pinion to the EPI arm. Compatible with (50W) vertical illuminators, DIC Optiphot 66 base EPI power receptical has 4 prongs, instead of 2 prongs on 1 June base. |
maintained by blekenbleu |